MF112_NTF

MF112_NTF - set of 10 SNOM probes, attached to tuning forks (quartz resonator on electrical plate holder), compatible with NT-MDT devices.

 

Material - single mode optical fiber Nufern 460HP.
Probe tip is coated by Al (70nm)/sublayer V (20nm) .
Tip diameter uncoated by Al (tip aperture)
70 - 130 nm (100±30)
Angle of fiber - about 20 degrees
Maximum optical input power - 400 microWatt
The probes are prepared by chemical etching method.
This method gives the optical efficiency 102 - 104 times better than those obtained by mechanical pulling.

 

 

Geometrical & Mechanical fiber  Specifications:

 

Clad Diameter

125.0 ± 1.5 µm

Coating Diameter

245 ± 15 µm

Core-Clad Concentricity

< 0.5 µm

Coating/Clad Offset

≤ 5 µm

Coating Material

UV Cured, Dual Acrylate

Operating Temperature

-55 to +85 °C

Short-Term Bend Radius

≥ 6 mm

Long-Term Bend Radius

≥ 13 mm

Proof Test Level

≥ 200 kpsi (1.4 GN/m²)

 

SNOM probe characteristics:

 

Characteristics

Probe type

MF112_NTF

Basic Nufern Fiber

460-HP

Operating wavelength, nm

450-600

Mode Field Diameter

3.5 ± 0.5 µm @ 515 nm

Second Mode Cut-Off, nm

430 ± 20

Optical efficiency for wavelength 530 nm

10-5 - 10-4

Optical efficiency for working wavelength

~ (10-5 - 10-4)

 

 

 

 

SNOM probe's end

Probe tip with Al coating.

Aperture is 70 - 130 nm (100±30).

 